PRINT if starting a new 4m mask use the new DIRECT_6958_4m_09Aug11.dat file ! L1=L2=4m, A1=20, A2=8 SANS/LARGE MASK/CLEAR MASK/CLEAR/TIME L/WAV 1.75 16.5 0.05/LOG ! strong scatter !L/Q .004 1. .08/LOG PRINT some WIDE Q bins for weak signals ! weak scatter L/Q 0.004, 0.001, 0.01, -0.15, 1.0 L/QXY 0 0.025 .005/lin ! note increase lower time limit here BACK/MON/TIMES 85000 98000 DET/REAR GRAVITY/ON mask h0 mask h190>h191 mask v0 mask v191 ! dead wire near top mask h156>h159 ! spot on rhs beam stop at 11m ! mask h57>h66+v134>v141 ! mask/time 13000 15250 !mask/time 5000 19200 ! L/R 41 -1 3 !L/SP 1 36864 !MON/LENGTH=17.937 2 ! ! 6 May direct beam improves overlap (based on polymer std) MON/DIRECT=DIRECT_RUN524_4m_06May11.dat MON/TRANS/SPECTRUM=2/INTERPOLATE ! set monitor 2 to interpolate ! PRINT scale not sorted yet !! MON/SPECTRUM=2/INTERPOLATE FIT/TRANS/LOG 1.75 16.5 ! ! 29/07/11 this centre found from gw20 9528 set centre 175.6 -168.8 ! ! 29/07/11 gdw20 9528 set scales 0.1845 1.0 1.0 1.0 1.0 ! correction to actual sample position, measured 115 before shutter, notionally 81mm before shutter SAMPLE/OFFSET -34 ! Correction to SANS2D encoders in mm DET/CORR REAR X -16.0 DET/CORR REAR Z 58.0 DET/CORR FRONT X -44.0 DET/CORR FRONT Y -20.0 DET/CORR FRONT Z 47.0 DET/CORR FRONT ROT 0.0 ! !! 29/07/11 MASKSANS2D.112H_4m_Lu stop flow (scale gdw with changer, offset above for stop flow)