1 | ! L1=L2=4m, A1=30x20, A2=12, large b/s |
---|
2 | MASK/CLEAR |
---|
3 | MASK/CLEAR/TIME |
---|
4 | L/WAV 1.75 16.5 0.05/LOG |
---|
5 | ! adjusted to suit weak scatter |
---|
6 | !L/Q 0.0045,0.001,0.0205,-.05,0.25,-.08,0.6 |
---|
7 | !L/Q 0.005,0.001,0.008,-0.08,1.0 |
---|
8 | ! strong scatter |
---|
9 | L/Q 0.004 1.0 0.08/log |
---|
10 | L/QXY 0 0.15 .002/lin |
---|
11 | ! note increase lower time limit here |
---|
12 | !BACK/MON/TIMES 85000 98000 |
---|
13 | BACK/M2 85000 98000 |
---|
14 | !check BACK/M4 |
---|
15 | DET/REAR |
---|
16 | GRAVITY/ON |
---|
17 | ! |
---|
18 | mask/rear h0 |
---|
19 | mask/rear h190>h191 |
---|
20 | mask/rear v0 |
---|
21 | mask/rear v191 |
---|
22 | mask/front h0 |
---|
23 | mask/front h190>h191 |
---|
24 | mask/front v0 |
---|
25 | mask/front v191 |
---|
26 | ! dead wire near top |
---|
27 | mask/front h156>h159 |
---|
28 | ! spot on top b/s when hit meniscus ??? |
---|
29 | !mask h72>h75+v117>v122 |
---|
30 | ! |
---|
31 | mask/time 13000 15250 |
---|
32 | !mask/time 5000 19200 |
---|
33 | PRINT masking off beamstop arm 12mm wide at 19 degrees |
---|
34 | !PRINT not masking beam stop arm, M4 out |
---|
35 | mask/line 13 19 |
---|
36 | ! |
---|
37 | L/R 47 -1 3 |
---|
38 | !L/SP 1 36864 |
---|
39 | !MON/LENGTH=17.937 2 |
---|
40 | ! |
---|
41 | ! use this for rear |
---|
42 | !MON/DIRECT=DIRECT_16246_4m_Rear_17Dec12.dat |
---|
43 | ! use for front, this command has to come second! |
---|
44 | !MON/DIRECT/FRONT=DIRECT_6958_4m_rescaled_09Mar12.dat |
---|
45 | MON/TRANS/SPECTRUM=2/INTERPOLATE |
---|
46 | TRANS/TRANSPEC=4/SHIFT=-70 |
---|
47 | !TRANS/TRANSPEC=3 |
---|
48 | ! set monitor 2 to interpolate |
---|
49 | MON/SPECTRUM=2/INTERPOLATE |
---|
50 | FIT/TRANS/LOG 1.75 16.5 |
---|
51 | !PRINT FIT/TRANS/OFF |
---|
52 | !FIT/TRANS/OFF |
---|
53 | ! |
---|
54 | ! 19/03/13 gdw20 17511 8mm beam |
---|
55 | set centre 173.0 -167.1 |
---|
56 | ! |
---|
57 | ! 19/03/13 gdw20 17511 8mm beam, 17513 8mm DTAB, 17514 12mm DTAB |
---|
58 | set scales 0.1925 1.0 1.0 1.0 1.0 |
---|
59 | ! correction to actual sample position, measured cryo at 320mm before shutter, notionally 81mm before shutter |
---|
60 | ! DLS set up 81 - 132 =-34 |
---|
61 | ! normal changer offset +53 |
---|
62 | SAMPLE/OFFSET 53 |
---|
63 | ! Correction to SANS2D encoders in mm |
---|
64 | DET/CORR REAR X -16.0 |
---|
65 | ! note rear Z was 58, but needs change all masks 9/11 on to 47 |
---|
66 | DET/CORR REAR Z 47.0 |
---|
67 | ! 7/8/12 front at 2.6m is at slightly too high Q values, was X -44 |
---|
68 | DET/CORR FRONT X -24.0 |
---|
69 | DET/CORR FRONT Y -20.0 |
---|
70 | DET/CORR FRONT Z 47.0 |
---|
71 | DET/CORR FRONT ROT 0.0 |
---|
72 | ! (cf 115D Perkins) 80 is just inside window, 386 instead of DT's 306, but seems to work? , 48 should be central grid |
---|
73 | DET/CORR FRONT RADIUS 48.0 |
---|
74 | DET/CORR FRONT SIDE 0.0 |
---|
75 | ! |
---|
76 | !! 19/03/13 MASK125O 4m A1=30x20,S5=20mm,S6=large,A2=12 Brooks |
---|