Ticket #6742 (closed: fixed)

Opened 8 years ago

Last modified 5 years ago

Investigate Calibration Failure of MAPS D4 Window

Reported by: Karl Palmen Owned by: Karl Palmen
Priority: major Milestone: Release 2.5
Component: Mantid Keywords:
Cc: anders.markvardsen@… Blocked By:
Blocking: Tester: Anders Markvardsen

Description

Investigate why the python tube calibration code cannot calibrate all or most of the tubes in the D4 window of the MAPS instrument for calibration run 14919, which is used by the MAPS example scripts.

If the cause is found and can be easily fixed, fix it, else report the situation.

Change History

comment:1 Changed 8 years ago by Karl Palmen

  • Status changed from new to assigned
  • Owner set to Karl Palmen

comment:2 Changed 8 years ago by Karl Palmen

  • Status changed from assigned to accepted

comment:3 Changed 8 years ago by Karl Palmen

I've found out that the failure of the calibration of the D4_window, is that the detector IDs for that window are in reverse order (step="-1") and the tube specification code expects them to be in numerical order. The IDLIST for the D4_window begins thus in the IDF file:

  <idlist idname="D4_window">
    <id start="44101256" step="-1" end="44101001" />

This can be easily fixed to work with step = -1, by checking the ID of the last detector. Also it could be fixed for step = any non-zero integer, but such a fix would be more complicated.

comment:4 Changed 8 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:5 Changed 8 years ago by Karl Palmen

To test, run the TubeCalibDemoMaps_D4.py example script in .../scripts/Calibration/Examples. Check the it does calibrate the tubes, even though they may be rotated a little (separate issue).

The branch is feature/6742_maps_d4window .

comment:6 Changed 8 years ago by Karl Palmen

Further investigation has shown that this simple solution has turned the tubes around so that the detectors of ID ...001 are in the right instead of the left. More work will need to be done.

comment:7 Changed 8 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:8 Changed 8 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:9 Changed 8 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:10 Changed 8 years ago by Karl Palmen

To test, run the TubeCalibDemoMaps_D4.py example script in .../scripts/Calibration/Examples. Check the it does calibrate the tubes and that the detector IDs do run from ...001 to ...256 left to right.

Branch = feature/6742_maps_d4window .

comment:11 Changed 8 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:12 Changed 8 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:13 Changed 8 years ago by Karl Palmen

  • Status changed from accepted to verify
  • Resolution set to fixed

comment:14 Changed 8 years ago by Anders Markvardsen

  • Status changed from verify to verifying
  • Tester set to Anders Markvardsen

comment:15 Changed 8 years ago by Anders Markvardsen

  • Status changed from verifying to closed

Works as expected.

To help the tester please specify the folder where a calibration MAPS file can be found, which in this case is
isis\inst$\NDXMAPS\Instrument\data\cycle_09_5

The steps I did to make run the script was

  1. added
    isis\inst$\NDXMAPS\Instrument\data\cycle_09_5 to Manage User Directories
  1. on line 16 in TubeCalibDemoMaps_D4.py changed path = r"C:/Temp/" to path = ""

comment:16 Changed 8 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:17 Changed 8 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:18 Changed 8 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:19 Changed 8 years ago by Karl Palmen

"To help the tester please specify the folder where a calibration MAPS file can be found, which in this case is isis\inst$\NDXMAPS\Instrument\data\cycle_09_5" This will be done as part of #6752 .

comment:20 Changed 8 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:21 Changed 8 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:22 Changed 8 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:23 Changed 7 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:24 Changed 7 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:25 Changed 7 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:26 Changed 7 years ago by Karl Palmen

Allow detectors to be numbered in reverse order in a tube re #6742

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 6080073c0f07929dd1deceed17d648e25c23c91b

comment:27 Changed 7 years ago by Karl Palmen

Deal with detectors in reverse order without reversing tube re #6742

This also fixes the problem of the tubes sloping.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: 25e4a73b88bf361e960b477ae09329e010e1cc62

comment:28 Changed 7 years ago by Karl Palmen

Tidy up python code re #6742

This commit was needed only because I made the previous one without saving.

Signed-off-by: Karl Palmen <karl.palmen@…>

Changeset: b75be43c026f82f2ee3a87168f83757f3d3d1618

comment:29 Changed 5 years ago by Stuart Campbell

This ticket has been transferred to github issue 7588