Ticket #9881 (closed: fixed)
EQSANS: write out transmission as output file
Reported by: | Mathieu Doucet | Owned by: | Mathieu Doucet |
---|---|---|---|
Priority: | major | Milestone: | Release 3.3 |
Component: | SANS | Keywords: | |
Cc: | Blocked By: | ||
Blocking: | Tester: | Vickie Lynch |
Description
The EQSANS team would like to have the raw and fitted transmission written to a file.
Change History
comment:2 Changed 6 years ago by Mathieu Doucet
Re #9881 save transmission to file
Changeset: 2479ff66322e9d713cf4515d4607f6bf15717015
comment:4 Changed 6 years ago by Mathieu Doucet
- Status changed from assigned to verify
- Resolution set to fixed
To test:
- Non frame-skipping mode: Start the ORNL SANS UI and reduce run 1466 from the system tests data directory. For the transmission, you can put the same run for the scattering data file, the sample direct beam file and the empty direct beam file. The transmission output will be 1.0. After the reduction, two transmission files should be written to disk: one for the transmission data, and one for the fit to that data. The files should appear in the directory that you chose on the first tab of the reduction UI as your output directory.
- Frame-skipping mode: Do the same with run 3293 from the system tests data directory. Two transmission files should also be written to disk.
Frame-skipping mode and non-frame-skipping mode will both produce two transmission files, but the processing goes through a different path in the code. Performing both tests ensures that the save method is correctly called in both paths.
comment:5 Changed 6 years ago by Vickie Lynch
- Status changed from verify to verifying
- Tester set to Vickie Lynch
comment:6 Changed 6 years ago by Vickie Lynch
- Status changed from verifying to closed
Merge remote branch 'origin/feature/9881_transmission_output_file'
Full changeset: 85a6cbebb04801402fd920cd8a557e2be18c7b9b
Note: See